STATEMENT OF WORK

1. SCOPE

1.1. Title: This Statement of Work defines the requirements for the inspection, scheduled annual preventative maintenance (PM), calibration, and unscheduled corrective repairs for the microscope identified in paragraph 1.3.
1.2. Location of Work: Portsmouth Naval Shipyard, Kittery, Maine. 1.3. Identification: Quanta 650 Electron Microscope (S/N: 9923437).

2. REFERENCES

2.1. American Society of Mechanical Engineers (ASME) B30 2.2. American National Standards Institute (ANSI)/American Gear Manufacturers Association

(AGMA) 2001-D04

2.3. American National Standards Institute (ANSI)/American Gear Manufacturers Association

(AGMA) 915-3-A99

2.4. OSHA 29 CFR 1910: Occupational Safety and Health Standards 2.5. Thermo Fisher Scientific/FEI Quanta 650 Instruction Manual (Rev. C, 2020): Standard operating and maintenance instructions for the identified microscope.

3. REQUIREMENTS

3.1. Unless indicated otherwise in this Statement of Work, the contractor will provide all personnel, equipment, tools, materials, supervision, and non-personal services necessary to perform comprehensive inspections, scheduled preventive maintenance, and unscheduled repairs for the microscope system.
3.2. The Contractor shall utilize only OEM-certified field service engineers who possess documented technical training specific to the identified Quanta 650 microscope platform. The Contractor shall be an OEM authorized service technician. 3.3. The Contractor shall comply with all federal, state, and PNSY safety and environmental regulations, including lock out/tag out procedures, electrical safety standards under OSHA 29 CFR 1910, and radiation safety guidelines associated with high-voltage electron column operation. The Contractor shall provide and require their personnel to wear personal protection equipment (PPE) in accordance with OSHA CFR 1926 and CFR 1915. 3.4. The Contractor shall use only new, OEM-approved parts, apertures, filaments, and lubricants for all maintenance and repair actions.
3.5. Inspection Services: The Contractor shall perform annual comprehensive physical and operational inspection of the microscope system. The Contractor shall: Spec #: 20XX-XX
Rev: -

3.5.1. Verify the integrity of the column vacuum system and measure vacuum leak rates. 3.5.2. Evaluate the electrical and thermal stability of the Schottky Field Emission Gun (FEG) source.
3.5.3. Inspect all imaging and analytical detectors (including ETD, BSE, and LFD) for sensitivity and noise level compliance.

The beginning of the document. Download the full file to read on.