Attachment 3 -.10-.Bldg 24 Pharmacy USP Testing Appendix
Updated
A PDF document (2.0 MB) attached to VACCHCS Pharmacy Certification, from the Department of Veterans Affairs. Due Nov 2, 2026, 10:00 AM (UTC-8).
File name: Attachment 3 - VFR2627401A-1.10-11 Sep 2026.Bldg 24 Pharmacy USP Testing.Appendix.pdf
The full file is for paid members, or buy this bid alone.
Preview
Project ID: NC-VFR2627401A-1
Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing
Appendix
Ver. 15 Oct 2026
4225 Executive Square, Suite 370
La Jolla, CA 92037
(800) 877-7742
Biological Safety Cabinet Certification Report
Cert.: DV2691010465-02
SO: NC-VFR2627401A-1
Contact: Travis Horn
Phone: (559) 225-6100 Ext.5735
Email: [email protected]
Facility: VA Fresno
Addr: 2615 East Clinton Avenue
Fresno, CA 93703
Bld: 24
Rm: HD Sterile Buffer Room
ID: 5505820
Cust ID: NA
Make: NUAIRE
Model: NU-540-400
SN: 199716071620
Class: II
Type: A2
Test Standard(s): MANUFACTURER, OSHA, ANSI/NSF 49, USP <797>, CETA CAG-003
Inspected
Calculated Inflow Velocity (FPM)
Average Supply Velocity Zone1 (FPM)
Average Supply Velocity Zone2 (FPM)
Average Supply Velocity Zone3 (FPM)
Average Supply Velocity Zone4 (FPM)
Supply Uniformity
Downflow Smoke Pattern Test
View Screen Retention Smoke Pattern Test
Work Opening Edge Retention Smoke Pattern Test
Sash Seal Smoke Pattern Test
Site Assessment
At-Rest Unidirectional Airflow Visualization Test
Dynamic Unidirectional Airflow Visualization Test
Supply Delta Pressure ("W.C.)
Duct Pressure ("W.C.)
Min
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
Max
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
Measured
NA
NA
NA
Pass
Pass
Pass
Pass
Pass
Pass
NA
Pass
NA
NA
Result
Pass
Pass
NA
NA
NA
Pass
Pass
Pass
Pass
Pass
Pass
NA
Pass
NA
NA
Supply
Min
Max
Measured
Results
HEPA As Found Max Point Leak (%)
0.000
0.010
0.000
Pass
HEPA As Left Max Point Leak (%)
0.000
0.010
NA
NA
Pre:0.00/New:0.00/Tot:0.00
Pass
Patch (%)
Aerosol Concentration (ug/l)
Pass
Exhaust
Min
Max
Measured
Results
HEPA As Found Max Point Leak (%)
0.000
0.010
0.000
Pass
HEPA As Left Max Point Leak (%)
0.000
0.010
NA
NA
Patch (%)
Pre:0.00/New:0.00/Tot:0.00
Pass
Aerosol Concentration (ug/l)
Pass
Comments: This certificate, DV2691010465-02, is supplement to test report DV2691010465. Added technician comments. Fabian Olivares 9/16/2026 12:30:40 PM
-------------------The space between the face of the BSC and permanent fixture directly opposite the access opening was less than 42 inches, used manufacturers alternative method to measure inflow readings.
**** Unit Certified ****
The following NIST-Traceable equipment were used to perform this test: Equipment ID Manufacturer
Model
Serial No.
The beginning of the document. Download the full file to read on.