Home / Fit Test Machine / Document

Salient Characteristics

Updated

A PDF document (118 KB) attached to Fit Test Machine, from the Department of the Air Force. Due Nov 18, 2026, 12:00 PM (UTC-6).

File name: Salient Characteristics.pdf

Download the full documentSee the bid

The full file is for paid members, or buy this bid alone.

Preview

Salient Characteristics
Name Brand: PortaCount Model 8040 Respirator Fit Tester

1. Purpose

The purpose of this letter is to outline the salient characteristics required for the procurement of a quantitative respirator fit tester (QNFT) to support mission-critical respiratory protection programs. The specified requirements ensure that the equipment provides accurate, objective, and automated fit testing for service members and personnel utilizing high-efficiency respiratory protection.

2. Technical Specifications & Performance

The following characteristics are mandatory for any equipment provided under this requirement to be considered technically acceptable:
Category

Characteristic

Mandatory Requirement

Technology

Detection Method

Must utilize Condensation Nuclei Counter (CNC)
technology to measure ambient aerosol concentrations
inside and outside the mask simultaneously.

Compatibility

Mask Types

Must be capable of testing all tight-fitting respirators with ≥99% filter efficiency, including SCBA, CBRN masks
(e.g., M50/M51), and elastomeric half/full facepieces.

Performance

Fit Factor Range

Must be capable of measuring Fit Factors from 1 to
10,000+.

Performance

Particle Range

Must detect particles in the 0.02 to 1.0 micrometer size range.

Category

Characteristic

Mandatory Requirement

Compliance

Standards

Must include pre-programmed automated protocols
compliant with OSHA 29 CFR 1910.134 and ANSI/ASSP
Z88.10.

3. Operational & Functional Requirements

Feature

Requirement

Real-Time Data

Equipment must feature a Real-Time FitCheck Mode to allow for immediate adjustment of the respirator during the donning process.

User Interface

Must feature a touchscreen interface for standalone operation and be compatible with PC-based software for group data management.

Connectivity

Must support data transfer via WiFi, Bluetooth, and USB-C.

Portability

The unit must be field-deployable, weighing no more than 5.0 lbs, and include a ruggedized transport case.

Group Testing

Software must allow a single administrator to manage up to four (4) units simultaneously from one interface.