RFQ 80NSSC27944070Q Q&A
1. For "or equal" o ers, will a system that covers the full 14,700-350 cm-1 range with automated MIR/NIR switching on one platform be accepted if its switching mechanism di ers in design (for example, the beamsplitter changer approach), provided all performance requirements are met? A di erent switching mechanism may be considered acceptable, provided that no repositioning or movement of the sample is required during operation and that all stated performance, functionality, and spectral range requirements are fully met. 2. Is the 30-days-after-receipt-of-order delivery firm, or will a longer manufacturer lead time be considered?
The stated delivery timeframe is preferred; however, longer manufacturer lead times may be considered as part of the evaluation of an o er.
3. Is on-site installation and training required? If so, please confirm the delivery and installation location at NASA Langley.
On-site installation and training are not required. The vendor shall provide detailed installation instructions su icient for Government personnel to install and configure the system. If a vendor elects to include on-site training and installation services, the installation location would be:
NASA Langley Research Center
Building 1146
11 W. Taylor St.
Hampton, VA 23681